Magnetic Inductive Flowmeters are one of the ideal flowmeter for liquids.
However, the measuring liquids have one condition;
The liquid to be measured is to be conductive..
Theoretically, the accuracy of Magnetic Inductive Flowmeters is free from
the difference and changes of liquid conductivity subject to the fact that
the such liquid conductivity is separate than the Minimum Limit of the
specification.
This Minimum Limit of the conductivity is specified by two different types
of liquid;
Water and Other liquids than Water
This is because that Flow noise is easily induced by water and the indication
and output are noisy and higher limit of conductivity is specified in case
of Water flow measurement.
Such Minimum Conductivity by types of flowmeter as well as types of liquids
is shown below;
| Version |
Model |
Liquids other than water |
Water |
Separate Version
  |
MAGMAX
MGS1000F+MGC110F
MGS4000F+MGC110F
MGS4500F+MGC110F
MGS5000F+MGC110F |
More than 5μS/cm |
More than 20μS/cm |
Compact (Integrated) Version
 |
MAGMAX
MGM1010K
MGM4010K
MGM4510K
MGM5010K
MGM1090K
MGM4090K
MGM4590K
MGM5090K |
More than 5μS/cm (More than 10μS/cm for DN2.5) |
More than 20μS/cm |
Capacitance detection type
 |
MAGMAX
MGM5090K-CAP |
More than 0.05μS/cm |
More than 1μS/cm |
|
Reference
| · |
Examples of liquids not suitable for Electro magnetic flowmeters due to
conductivity; |
| |
|
Petroleum, Petroleum products, Liquefied gases, Vegetable oil, Pure and
Ultrapure water, Pure Alcohol |
| · |
Seems to be OK, but failed |
| |
|
Spirit having 25% alcohol. Whiskey just after distilled, 30% Propylene
mixed water |
| · |
Seems to be difficult, but successfully measured; |
| |
|
Shampoo |
| · |
Successful case by using Capacitance Detection type model MGM5090K-CAP |
| |
|
Sugar syrup, Ethylene Glycol, Boiler water
(They could not be measured by existing Magnetic flowmeters.) |
|