To obtain specified accuracy, straight run both for upstream and downstream is needed. The required straight run is depding on the type of flowmeter.
The recommended straight run by types is shown below.

Type of Flowmeter Tokyo Keiso Model Required straight run Remarks
Upstream Downstream
Variable Area Flowmeter R-100
AC
AM-1000
M
A
P etc.
Not required Not required Turbulence in large size flowmeter may result unstable float movementr.
Magnetic Flowmeter MAGMAX Series 5D* 2D Flow control valves are to be located downstream, if any.
Ultrasonic Flowmeter UCUF Not required Not required Flow control valves are to be located downstream, if any.
UL600N
UL610P
UL500K
10D 5D
DP type Flowmeter Orifice type O-100
O-1000
O-180(780)
20D 10D ---
Pitot Tube type HDT
CFW-1000, 2000
CDT
---
V Cone Flowmeter VC,VD,VM Not required 3D 2D upstream and 5D downstream is recommended in case that valve and other fittings than elbows are located.
Vortex Flowmeter VF-1000
VF-2000
VF-2200
VF-3000
VF-4000
KARUMAN ACE
10D 5D Flow control valves are to be located downstream, if any.
Wheel type Flowmeter W-100
W-200
W-300
W-400
W-800
Not required Not required 5D both upstream and downstream for type
TF-1150,A1250,A1251,A1551,A2251.
5D upstream 0D for downstream for
TF-2421~2441.
W-500
WX-2000
20D 5D Flow control valves are to be located downstream, if any.
Corioli Force Mass flowmeter MASSMAX7000
MASSMAX7100
Not required Not required 5D both upstream and downstream for type
TF-1150,A1250,A1251,A1551,A2251.
5D upstream 0D for downstream for
TF-2421~2441.
Thermal Mass flowmeter/Conroller TF
TC
HM
Not required Not required 5D both upstream and downstream for type
TF-1150,A1250,A1251,A1551,A2251.
5D upstream 0D for downstream for
TF-2421~2441.
Thermal flowmeter TH 20D* 5D 3D upstream 0D for downstream for TH-1600 and TH-1800 with flow strainer.

D: Inside diameter of piping. (5D: 5 times of Inside diameter of piping)

Recommended straight run in the table is subject to the condition that one 90 deg elbow is located upstream. Double elbows, pumps and valves upstream may require further straight run.